Loadlock

The loadlocks for both endstations have 6-slot carousels. While heating was originally possible in the loadlock, in 2023 the carousel was replaced with a simpler design that does a much better job at not dropping samples, but does not permit sample heating.

The carousel can be continuously rotated with a manual knob at the bottom of the chamber:

_images/carousel.jpeg

On the A endstation, venting or pumping the loadlock is a one-button task from the ‘ARPES Control panel’ window on the endstation PC

_images/interface.png

It typically happens when venting that the N2 inlet is closed slightly too early and the loadlock remains in a weak vacuum. In this case you can either gently pull/pry a little on the door to get it open, OR manually re-enable the N2 venting. Right click on the loadlock turbo pump icon in the control panel to gain buttons for this. Don’t forget to disable ventilation again before trying to start the pump down macro.

_images/manual_vent.png

A special tool (shown below) is required to insert and remove samples via the viton-sealed door. It typically lives in the BLOCH workshop room. A staff member will show you how to operate it.

_images/tool.png

Pumpdown time

Pressure (mBar)

Fast turnaround (hh:mm)

Slow turnaround (hh:mm)

1x10-6

00:15

00:30

1x10-7

2:00

3:50

1x10-8

22:00

??

Measured February 2019. Quick means only the amount of time necessary to insert a sample. Slow means opened, then left vented with door pushed closed for 30mins

Base pressure without baking: 6x10-9 mBar

We typically recommend waiting for a pressure of 5x10-7 mBar or better before opening to the transfer chamber, which you should obtain within 1-2 hours.