LEED
Secondary Preparation Chamber
Prep Sec has an Omicron SpectaLEED, with the gun swapped out for an OCI unit. It is controlled entirely through Omicron’s LEED control software, installed on the endstation laptop. We sincerely apologize for making you use this software. Note in particular:
- The software will not run unless:
The control electronics are switched on (power supply box at the bottom of the rack with an orange power button at the front)
The USB cable from the control box was the first USB cable plugged into the computer.
The uEYE LEED camera is plugged into the computer
If the software is closed or the laptop lid is closed, the filament current is immediately dumped to zero - please avoid this.
Eventually it will stop accepting any keyboard entry. Values can only be changed using the mouse, or by copy-pasting in values you have written into a text editor. You can fix this by restarting the program, but it will require a slow filament ramp down and back up, see above.
By default the filament current is locked to 1.7A. To get the extra 50-100mA you sometimes need, switch to the higher current range.
To save an image, right click on it and choose ‘save snapshot’. It will include all metadata about the settings in the image.
When working at beam energies below 50eV it is important to bring the LEED with its mu-metal shield as close as possible to the sample. Don’t forget to retract again when you’ve finished before moving the manipulator away, or there may be a collision.
Recommended settings
Set the camera to 1024x768, 20MHz.
Filament current |
1.83 A |
Wehnhelt |
1.5 |
L2 Offset |
5 |
L2 Gain |
0.392 |
L1/3 Offset |
8.4 |
L1/3 Gain |
1.54 |
Screen HV |
5kV |
Main Preparation Chamber
Prep Main also has a SpectaLEED, but with the original gun. The operating current is lower, only 1.4A. Here the LEED is controlled with the older style analog control box sitting at the top of the rack, but the LEED software is still needed to interface with the camera.
When working at beam energies below 50eV it is important to bring the LEED with its mu-metal shield as close as possible to the sample. Don’t forget to retract again when you’ve finished before moving the manipulator away, or there may be a collision.