Alkali deposition ================================= The analyzer chamber has a permanently mounted cluster alkali source (SAES type), enabling deposition of K, Cs and Rb on the measurement manipulator. This is done at the upper transfer level of the chamber - it's not possible to deposit while measuring. .. image:: /_static/alkali/clustersource.jpg :width: 500 :align: center | ===================== ======================== ======================== ================================ Source SAES product code Evaporation starts Suggested core level to monitor ===================== ======================== ======================== ================================ Rb RB/NF/4.8/17 FT10+10 5.3 A Rb 3d at 111.5eV K K/NF/3.1/17 FT10+10 5.3 A K 3p at ~18.5eV Cs CS/NF/5.5/17 FT10+10 4.7 A Cs 4d at 77eV ===================== ======================== ======================== ================================ See also: :download:`SAES datasheet <_static/alkali/datasheet.pdf>` All sources can be degassed up to 3.3A without anything happening. You will likely need to perform multiple depositions to get the coverage you want. A recommended starting current is 5.6A for 60s, then scale the time or current up or down depending on how that goes. This corresponds to around 700\ |deg|\ C, so the source will be glowing: .. image:: /_static/alkali/glowing.jpg :width: 500 :align: center | However note that the evaporation you will achieve is a non-linear function of current and of time. The sticking coefficient is also highly dependent on the sample surface and temperature. Some samples require evaporation cycles of only a few seconds where others might need minutes. As a general rule, alkali evaporation will not work above 200K due to rapid re-evaporation or intercalation. To operate the source, connect a DC power supply with the custom cable set. Please be very gentle with the feedthrough pins! .. image:: /_static/alkali/powersupply.png :align: center | Below are two example measurements taken on SiC graphene while testing the Rb source. At hv=180eV there is already a small Rb core level present after 30s / 5.5A, and the doping had not saturated after 5 deposition cycles of escalating time and current: .. image:: /_static/alkali/shortDose.png :align: center | .. image:: /_static/alkali/longDose.png :align: center | .. |deg| unicode:: U+00B0